FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 2982865 · Received January 31, 2013

Report

Report Number
1220908-2013-00123
Event Type
Malfunction
Date Received
January 31, 2013
Report Date
January 10, 2013
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K060559
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED MALFUNCTION WAS OBSERVED AND ATTRIBUTED TO A FAULTY SOLDER JOINT ON THE PACE/DEFIB ENGINE BOARD. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43613 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA