FDA Adverse Event Malfunction Summary report: N

POSEY KEEPSAFE DELUXE

MDR report key: 2982840 · Received January 31, 2013

Report

Report Number
2020362-2013-00039
Event Type
Malfunction
Date Received
January 31, 2013
Report Date
January 2, 2013
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE; THE ALARM DID NOT POWER ON WHEN USING NEW BATTERIES OR AN EXTERNAL POWER SUPPLY AND THE BATTERY DOOR IS MISSING. THE ALARM ONLY POWERS UP WHEN USING A 9V ADAPTER AND PASSED ALL FUNCTIONAL TESTS. THE BATTERY SPRINGS HAVE BATTERY LEAKAGE RESIDUE ON THEM AND ARE CORRODED. THERE IS BATTERY LEAKAGE RESIDUE INSIDE THE BATTERY COMPARTMENT. NOTE: INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: VISUALLY INSPECT THE UNIT FOR CRACKS IN THE CASE, MISSING SCREWS, MISSING BATTERY DOOR, BROKEN WIRES OR SIGNS OF EXPOSURE TO LIQUID. AFTER EACH INCIDENT, YOU MUST VERIFY THAT THE UNIT IS WORKING PROPERLY. CHECK TO MAKE SURE THE AUDIBLE ALARM IS FUNCTIONING PROPERLY BY APPLYING AND LIFTING WEIGHT OFF SENSOR IN SEVERAL SPOTS TO ACTIVATE THE ALARM. DO NOT USE THE POSEY KEEPSAFE DELUXE IF THE AUDIBLE ALARM DOES NOT SOUND. TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. ALWAYS INSTALLED A COMPLETELY NEW SET OF BATTERIES WHEN THE CHIRPING DUE TO LOW BATTERIES POWER SOUND BEGINS. DO NOT REPLACE A SINGLE CELL, BUT ALL CELLS IN THE ALARM. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK (4 BATTERIES). USE OF MIXED BATTERIES OR BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. REMOVE ANY ALARM FROM USE AND SEND TO THE APPROPRIATE FACILITY AUTHORITY IF BATTERIES ARE DAMAGED OR CORRODED OR THE BATTERY COMPARTMENT HAS SIGNS OF PREVIOUS BATTERY CORROSION SUCH AS WHITE POWDER RESIDUE. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THE ALARM WILL NOT POWER ON. THE CUSTOMER REPORTED THEY HAVE REPLACED THE BATTERIES WITH NEW SUPPLY AND THE ALARM MAYBE MISSING THE BATTERY COVER. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41928 POSEY KEEPSAFE DELUXE KMI J. T. POSEY CO. 8374 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK OVER-THE-MATTRESS SENSOR PAD MODEL: 8308, LOT #UNK