FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2982795 · Received February 28, 2013

Report

Report Number
1416980-2013-04906
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 9, 2013
Report Date
February 10, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT WAS CONFIRMED. THE CAUSE OF THE EVENT WAS USE ERROR. THERE WAS NO REPORTED DEVICE MALFUNCTION THEREFORE NO FURTHER INVESTIGATION WILL BE PERFORMED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A CHECK HEATER LINE ALARM WHICH OCCURRED ON THE HOMECHOICE (HC) DURING FILL 1 OF 4, AND DURING TROUBLESHOOTING A LEAK WAS FOUND. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) CONFIRMED WITH THE HOME PATIENT (HP) THAT THE FRANGIBLES HAD BEEN BROKEN. THE HP INSPECTED THE DRAIN LINE, MOVED THE RED CLAMP THREE INCHES IN EITHER DIRECTION, AND INSPECTED THE HEATER BAG PORT. THE HP THEN MENTIONED THAT THERE WAS SOME FLUID ON HER NIGHT STAND. THE TSR ASSISTED THE HP WITH ENDING THERAPY DUE TO THE POSSIBLE RISK OF THE SET BEING COMPROMISED. THE TSR ADVISED THE HP TO DISPOSE OF THE CURRENT SUPPLIES AND START OVER WITH ALL NEW SUPPLIES. THE HC WAS OPERATIONAL. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. BAXTER PRODUCT SURVEILLANCE SPOKE WITH THE HP REGARDING THE EVENT. SHE STATED THAT THE LEAK WAS COMING FROM THE CONNECTION BETWEEN A SUPPLY BAG AND THE CASSETTE. SHE STATED THAT SHE HAD NOT TIGHTENED THE CONNECTION ALL OF THE WAY, AND THERE WERE NO DEFECTS NOTED ON THE SUPPLIES. THERAPY SINCE HAS BEEN GOING WELL, AND THERE WERE NO ADVERSE EFFECTS REPORTED IN RELATION TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87574 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 68 YR HOMECHOICE