FDA Adverse Event Malfunction Summary report: N

MUSTANG¿

MDR report key: 2982763 · Received February 28, 2013

Report

Report Number
2134265-2013-01122
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 5, 2013
Report Date
February 5, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LIT
PMA / PMN Number
K103751
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT VASCULAR ACCESS WAS OBTAINED VIA THE RIGHT FEMORAL ARTERY. THE 8MMX80MM EPIC STENT REMAINED WELL POSITIONED AND WELL APPOSED TO THE VESSEL WALL. ONE PRE DILATION WAS PERFORMED WITH A 5MM-40MM NON BSC BALLOON CATHETER AT 16ATMS. THE 7.0 X 40, 135CM MUSTANG BALLOON CATHETER WAS REMOVED INTACT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS LEFT EXTERNAL ILIAC ARTERY (EIA). THE TARGET LESION WAS PREDILATED WITH A 5MMX40MM BALLOON CATHETER. A 8MMX80MM EPIC STENT WAS IMPLANTED. FOR POST DILATION, DURING THE FIRST INFLATION OF A 7.0 X 40, 135CM MUSTANG BALLOON CATHETER AT 16ATMS A BALLOON RUPTURE OCCURRED. THE MUSTANG BALLOON CATHETER WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS LISTED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87099 MUSTANG¿ CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - GALWAY H74939171070410 15626909

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE, V-18 300CM| INTRODUCER SHEATH, TERUMO 6FR DESTINATION