FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 2982744 · Received February 25, 2013

Report

Report Number
1419937-2013-00121
Event Type
Malfunction
Date Received
February 25, 2013
Date of Event
January 23, 2013
Report Date
January 23, 2013
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER WAS SENT A REPLACEMENT POWER SUPPLY. CONTACT WAS NOT MADE WITH THE CUSTOMER TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVAL/INVESTIGATION AT THIS TIME. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. HOWEVER, THE CUSTOMER STATED THE POWER CORD BURNT APART, WHICH COULD BE A SAFETY RISK. THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW-UP REPORT WILL BE FILED AT THAT TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE POWER SUPPLY FOR HER BREAST PUMP GOT HOT AND "BURNED APART."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
79179 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063 / 9207010 REV L

Patients

Seq Age Sex Outcome Treatment
1