FDA Adverse Event Malfunction Summary report: N

OSTEO-WEDGE OPENING WEDGE BONE LOCKING SYSTEM

MDR report key: 2982669 · Received February 25, 2013

Report

Report Number
3004993707-2013-00002
Event Type
Malfunction
Date Received
February 25, 2013
Date of Event
June 25, 2012
Report Date
February 19, 2013
Manufacturer
RAHAM MEDICAL TECHNOLOGIES LLC (DBA GRAMEDICA)
Product Code
HRS
PMA / PMN Number
K111326
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONLY A PORTION OF THE SCREW WAS AVAILABLE FOR EVALUATION. ONLY A PORTION OF THE BROKEN SCREW COULD BE EVALUATED. EVALUATION SUMMARY: TESTING WAS NOT PERFORMED ON THE SCREW, ONLY A PORTION OF THE SCREW WAS RETURNED. A VISUAL INSPECTION WAS PERFORMED ON THE RETURNED PORTION OF THE SCREW USING 5X MAGNIFICATION AND CALIPERS AND REVEALED THAT BREAK OCCURRED AT .221/.241" (ANGLED) FROM THE TOP OF THE HEAD. THE VISUAL INSPECTION DID NOT REVEAL A KNOWN CAUSE FOR THE SCREW TO BREAK. A REVIEW OF THE MANUFACTURING PAPERWORK (DEVICE RECORDS) REVEALED THAT THE PROCESSES WERE MET ACCORDING TO DRAWING SPECIFICATIONS HELD IN THE DEVICE MASTER RECORD. A REVIEW OF OUR INSPECTION PAPERWORK ALSO CONFIRMED THAT THE MFG. PROCESSES WERE MET ACCORDING THE DEVICE MASTER RECORD. AN INCOMPLETE DEVICE WAS RETURNED AND THE CAUSE OF THE FAILURE WAS NOT FOUND. GRAMEDICA RECEIVED THE DEVICE IN A CONDITION MAKING A THOROUGH EVALUATION IMPOSSIBLE. FOR METALLIC INTERNAL FIXATIONS DEVICES THIS IS A KNOWN POTENTIAL RISK, AND STATED ON OUR PACKAGE INSERT, INSTRUCTIONS FOR USE REV. 1. GRAMEDICA HAS REDUCED THIS POTENTIAL RISK BY PERFORMING MECHANICAL TESTS ON THE SCREWS PER (B)(4) FROM A MEDIUM RISK LEVEL TO A LOW RISK LEVEL.

Description of Event or Problem · 1

PATIENT PRESENTED WITH PERSISTENT/CONSTANT PAIN, LEADING TO THE DISCOVERY OF THE BROKEN SCREW. REMOVAL WAS PERFORMED ON (B)(6) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80771 OSTEO-WEDGE OPENING WEDGE BONE LOCKING SYSTEM BONE LOCKING PLATE SYSTEM HRS RAHAM MEDICAL TECHNOLOGIES LLC (DBA GRAMEDICA) GM-200-14 14656-030

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention