FDA Adverse Event Other Summary report: N

NUCLETRON MICROSELCTRON

MDR report key: 2982480 · Received October 18, 2007

Report

Report Number
9611894-2006-00002
Event Type
Other
Date Received
October 18, 2007
Date of Event
September 26, 2006
Report Date
October 16, 2006
Manufacturer
NUCLETRON B.V.
Product Code
JAQ
PMA / PMN Number
K953946
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PROBLEM WITH THE TRANSFER TUBES WAS IDENTIFIED DURING PRODUCTION. USER FACILITIES HAVE NOT REPORTED PROBLEMS RELATED TO AN INCREASE IN LENGTH OF THE TRANSFER TUBE. SEE SCANNED PAGE.

Description of Event or Problem · 1

THE CRIMPING OF THE TRANSFER TUBE COUPLINGS HAS FOUND TO BE INADEQUATE ON RECENTLY MANUFACTURED TRANSFER TUBES. WITH INADEQUATE CRIMPING, THE COUPLING CAN SLIDE OVER THE TRANSFER TUBE MATERIAL. AS A RESULT, THE TRANSFER TUBE CAN ELONGATE. IF FORCE IS APPLIED, THE OVERALL LENGTH OF THE TRANSFER TUBE COULD INCREASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLETRON MICROSELCTRON JAQ REMOTE AFTERLOADER ACCESSRY JAQ NUCLETRON B.V. 009979 079293-02

Patients

Seq Age Sex Outcome Treatment
1