FDA Adverse Event Malfunction Summary report: N

SECURE II MED/SURG BED

MDR report key: 2982395 · Received February 28, 2013

Report

Report Number
0001831750-2013-01580
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 4, 2013
Report Date
February 4, 2013
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE BRAKES WOULD NOT DISENGAGE DUE TO BENT HEAD END BRAKE CAM. THE BRAKES COULD BE MANUALLY DISENGAGED, BUT IT WOULD REQUIRE THAT THE BRAKE PEDAL BE LIFTED, WHICH MAY NOT BE INTUITIVE TO THE USER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.