6.0MMTI PANGEA (TM) POLYAXIAL
Report
- Report Number
- 2530088-2013-10106
- Event Type
- Injury
- Date Received
- February 26, 2013
- Report Date
- February 23, 2011
- Manufacturer
- SYNTHES BRANDYWINE
- Product Code
- NKB
- PMA / PMN Number
- K103287
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DHR REVIEW WAS PERFORMED: A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR MANUFACTURING REVEALED NO COMPLAINT RELATED ISSUES. POLYAXIAL ASSEMBLY COMPOSITION WAS CONFIRMED USING A MATERIAL ANALYZER. HOWEVER NOT ALL THE INDIVIDUAL COMPONENTS COULD BE ANALYZED SEPARATELY. BASED ON DHR REVIEW AND INVESTIGATION RESULTS THIS COMPLAINT IS INDETERMINATE.
IT WAS REPORTED PATIENT UNDERWENT L3 TO S1 POSTERIOR INSTRUMENTED FUSION USING SYNTHES PANGEA PEDICLE SCREW SYSTEM TO TREAT RECURRENT SPINAL STENOSIS IN THE SETTING OF POSTLAMINECTOMY SYNDROME AND POSTLAMINECTOMY INSTABILITY ON (B)(6) 2011. PATIENT DID WELL INITIALLY BUT STARTED TO COMPLAIN OF ONSET OF LOW BACK AND LEFT LEG PAIN 6 WEEKS FOLLOWING THE SPINAL FUSION. WORK UP SHOWED DISENGAGEMENT OF THE LEFT SIDED L1-S1 PEDICLE SCREWS WITH THE ROD MIGRATING ROSTRALLY IMPINGING ON THE L2-L3 FACET JOINT. LUMBAR CT SCAN AND LUMBAR FILMS SHOWED DISENGAGEMENT OF THE FUSION ROD FROM THE S1 PEDICLE SCREW AND ITS PROXIMAL MIGRATION. PATIENT UNDERWENT REVISION DECOMPRESSION AND INSTRUMENTED TRANSFORAMINAL LUMBAR INTERBODY FUSION FROM L3-S1 FOR SEGMENTAL INSTABILITY AND RECURRENT SPINAL STENOSIS. THIS IS 5 OF 19 RECORDS FOR THE SAME EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 81332 | 6.0MMTI PANGEA (TM) POLYAXIAL | NKB | SYNTHES BRANDYWINE | 6538752 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Required Intervention |