FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2982294 · Received February 28, 2013

Report

Report Number
1416980-2013-04887
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 15, 2013
Report Date
February 15, 2013
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING PREVENTATIVE MAINTENANCE. THE ROOT CAUSE OF THE F-2 ALARM WAS DETERMINED TO BE DAMAGE TO THE DOOR LATCH ROLLER. TO CORRECT THE CONDITION, THE DOOR LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED THE F-2 ALARM. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86387 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1