FDA Adverse Event Malfunction Summary report: N

DEVICE #4- SYNERGIES PLUS NC2

MDR report key: 2982223 · Received November 1, 2005

Report

Report Number
2919016-2005-00043
Event Type
Malfunction
Date Received
November 1, 2005
Date of Event
July 25, 2005
Report Date
October 3, 2005
Manufacturer
DADE BEHRING, INC.
Product Code
LON
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METHOD ¿ OTHER- TESTED ISOLATE ON MULTIPLE PANEL TYPES, WITH MULTIPLE INOCULATION METHODS. TESTED D-12 SPECIMEN ON FROZEN REFERENCE PANELS. ROUTINE MONITORING OF COMPLAINT HISTORY AND PERFORMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. RESULTS- OTHER- IN-HOUSE TESTING OF D-12 SPECIMEN YIELDED FALSE SUSCEPTIBLE RESULTS WHEN COMPARED TO FROZEN REFERENCE PANELS. COMPLAINT REVIEW INDICATES PRODUCT IS PERFORMING AS EXPECTED. CONCLUSION- OTHER ¿ UNUSUAL EVENT ¿ IN HOUSE TESTING OF D-12 SPECIMEN PROVIDED FALSE SUSCEPTIBLE RESULTS WHEN COMPARED TO FROZEN REFERENCE PANELS. OVERALL PRODUCT PERFORMANCE IS ACCEPTABLE. DEVICE 4 OF 4. REFERENCE MFR REPORT NUMBERS: 2919016-2005-0031, 00041, 0042.

Description of Event or Problem · 1

DADE BEHRING PARTICIPATED IN THE 2005 CAP BACTERIOLOGY SURVEY D AND OBTAINED LOWERED MICS FOR S. MALTOPHILIA D-12 SPECIMEN FOR CEFTAZIDIME IN INTERNAL TESTING. S. MALTOPHILIA YIELDED FALSE SUSCEPTIBLE RESULTS FOR CEFTAZIDIME (CAZ) WHEN COMPARED TO FROZEN REFERENCE PANELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEVICE #4- SYNERGIES PLUS NC2 ANTIMICROBIAL SUSCEPT. TEST PANEL: DRIED NEG LON DADE BEHRING, INC. NA 2005-09-24

Patients

Seq Age Sex Outcome Treatment
1