DEVICE #4- SYNERGIES PLUS NC2
Report
- Report Number
- 2919016-2005-00043
- Event Type
- Malfunction
- Date Received
- November 1, 2005
- Date of Event
- July 25, 2005
- Report Date
- October 3, 2005
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- LON
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
METHOD ¿ OTHER- TESTED ISOLATE ON MULTIPLE PANEL TYPES, WITH MULTIPLE INOCULATION METHODS. TESTED D-12 SPECIMEN ON FROZEN REFERENCE PANELS. ROUTINE MONITORING OF COMPLAINT HISTORY AND PERFORMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. RESULTS- OTHER- IN-HOUSE TESTING OF D-12 SPECIMEN YIELDED FALSE SUSCEPTIBLE RESULTS WHEN COMPARED TO FROZEN REFERENCE PANELS. COMPLAINT REVIEW INDICATES PRODUCT IS PERFORMING AS EXPECTED. CONCLUSION- OTHER ¿ UNUSUAL EVENT ¿ IN HOUSE TESTING OF D-12 SPECIMEN PROVIDED FALSE SUSCEPTIBLE RESULTS WHEN COMPARED TO FROZEN REFERENCE PANELS. OVERALL PRODUCT PERFORMANCE IS ACCEPTABLE. DEVICE 4 OF 4. REFERENCE MFR REPORT NUMBERS: 2919016-2005-0031, 00041, 0042.
DADE BEHRING PARTICIPATED IN THE 2005 CAP BACTERIOLOGY SURVEY D AND OBTAINED LOWERED MICS FOR S. MALTOPHILIA D-12 SPECIMEN FOR CEFTAZIDIME IN INTERNAL TESTING. S. MALTOPHILIA YIELDED FALSE SUSCEPTIBLE RESULTS FOR CEFTAZIDIME (CAZ) WHEN COMPARED TO FROZEN REFERENCE PANELS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEVICE #4- SYNERGIES PLUS NC2 | ANTIMICROBIAL SUSCEPT. TEST PANEL: DRIED NEG | LON | DADE BEHRING, INC. | NA | 2005-09-24 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |