FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2982104 · Received February 8, 2013

Report

Report Number
1720753-2013-01593
Event Type
Malfunction
Date Received
February 8, 2013
Date of Event
January 18, 2013
Report Date
February 8, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE SYSTEM WAS RESET DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK ITO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A COLLIMATOR IRIS POT ERROR THAT OCCURRED DURING A PROCEDURE WITH PT INVOLVEMENT, THEREFORE, THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ARO. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55179 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1