FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK INFUSION SET

MDR report key: 2982066 · Received February 8, 2013

Report

Report Number
2183996-2013-00194
Event Type
Malfunction
Date Received
February 8, 2013
Date of Event
January 29, 2013
Report Date
May 8, 2013
Manufacturer
ROCHE HEALTH SOLUTIONS INC.
Product Code
FPA
PMA / PMN Number
UNK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION THE COMPLAINT CANNOT BE VERIFIED DUE TO MISHANDLING OF THE PRODUCT BY THE CUSTOMER. RESULT TRANSFER SET: A VISUAL INSPECTION AND A TEST FOR FLOW AND LEAK WERE PERFORMED ON THE RETURNED USED SAMPLE. THE TUBING WAS SPLIT FROM THE RESERVOIR LUER CONNECTOR AND FOR THIS CONDITION WAS LEAKING. THE FLOW TEST WAS IN ACCORDANCE WITH SPECIFICATIONS. THE STATIC PULL TEST WAS NOT PERFORMED BECAUSE THE SAMPLES WILL BE USED FOR FURTHER ANALYSIS. THE MANUFACTURER TESTS ALL PRODUCTS DURING PRODUCTION FOR LEAK AND FLOW. THERE IS NO INDICATION THAT A NONSPECIFIC PRODUCT HAS BEEN DELIVERED TO ANY CUSTOMER. PUMP: AS THE PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION AND THE PRODUCTION DATA COMPLY WITH THE SPECIFICATION, THE COMPLAINT COULD NOT BE REPLICATED. PRODUCTION REPORTS WERE REVIEWED. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO MISHANDLING OF THE PRODUCT BY THE CUSTOMER.

Description of Event or Problem · 1

ON (B)(6) 2013, HUSBAND REPORTED THE INFUSION SET BROKE INTO TWO PIECES NEAR THE MIDDLE OF THE TUBE. THE INFUSION DEVICE WAS NOT IN USE AT THE TIME. HUSBAND WAS UNABLE TO PROVIDE THE TYPE, LOT, OR EXPIRATION DATE OF THE PRODUCT. HE BELIEVES THEY HAVE HAD THE PRODUCT FOR A "LONG TIME." HUSBAND WAS ADVISED TO DISCARD PRODUCT IF EXPIRED. THE INFUSION SET WAS REQUESTED FOR EVALUATION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND THE PATIENT DID NOT REQUIRE TREATMENT FROM A HEALTH CARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55176 ACCU-CHEK INFUSION SET FPA ROCHE HEALTH SOLUTIONS INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 78 YR INSULIN INFUSION PUMP| RELATED ACCESSORIES| INSULIN