CORE SUMEX DRILL
Report
- Report Number
- 0001811755-2013-00395
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- January 26, 2013
- Report Date
- January 31, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE FAILURE COULD NOT BE CONFIRMED BECAUSE THE REPORTED EVENT OF HEAT DUE TO A BIAS-CURRENT WARNING PREVENTING THE HANDPIECE FROM RUNNING. THE TECHNICIAN NOTED THAT THE ROTOR ASSEMBLY WAS CORRODED. BASED ON (B)(6) 2013 TRENDING, THE MOST LIKELY CAUSE FOR THE REPORTED EVENT WOULD BE THE CORRODED ROTOR ASSEMBLY.
THE DEVICE WAS RETURNED TO THE MANUFACTURER AND AN INVESTIGATION IS ANTICIPATED. A FOLLOW UP REPORT WILL BE SUBMITTED IF THE INVESTIGATION RESULTS REQUIRE.
IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY THE DEVICE OVERHEATED. HOWEVER DURING DEVICE EVALUATION, THE DEVICE DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAD THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.
IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY THE DEVICE OVERHEATED. HOWEVER DURING DEVICE EVALUATION, THE DEVICE DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAD THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83213 | CORE SUMEX DRILL | DRILL, SURGICAL, ENT | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |