FDA Adverse Event Malfunction Summary report: N

CORE SUMEX DRILL

MDR report key: 2981846 · Received February 27, 2013

Report

Report Number
0001811755-2013-00395
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
January 26, 2013
Report Date
January 31, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE FAILURE COULD NOT BE CONFIRMED BECAUSE THE REPORTED EVENT OF HEAT DUE TO A BIAS-CURRENT WARNING PREVENTING THE HANDPIECE FROM RUNNING. THE TECHNICIAN NOTED THAT THE ROTOR ASSEMBLY WAS CORRODED. BASED ON (B)(6) 2013 TRENDING, THE MOST LIKELY CAUSE FOR THE REPORTED EVENT WOULD BE THE CORRODED ROTOR ASSEMBLY.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AND AN INVESTIGATION IS ANTICIPATED.  A FOLLOW UP REPORT WILL BE SUBMITTED IF THE INVESTIGATION RESULTS REQUIRE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY THE DEVICE OVERHEATED. HOWEVER DURING DEVICE EVALUATION, THE DEVICE DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAD THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY THE DEVICE OVERHEATED. HOWEVER DURING DEVICE EVALUATION, THE DEVICE DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAD THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83213 CORE SUMEX DRILL DRILL, SURGICAL, ENT ERL STRYKER INSTRUMENTS-KALAMAZOO