FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 SYNCHRON® SYSTEM

MDR report key: 2981758 · Received February 27, 2013

Report

Report Number
2050012-2013-00124
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
January 30, 2013
Report Date
January 31, 2013
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED AS THE CUSTOMER RESOLVED THE ISSUE BY TIGHTENING THE CARTRIDGE CHEMISTRY SAMPLE SYRINGE. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED (B)(6) RESULTS, FOR TWO PATIENTS, INVOLVING THE UNICEL DXC 800 SYNCHRON SYSTEM. ALANINE AMINOTRANSFERASE (ALT), ASPARTINE AMINOTRANSFERASE (AST), TRIGLYCERIDES (TG) AND MAGNESIUM (MG) RESULTS WERE SUPPRESSED WITH BLANK ERRORS FOR AN UNKNOWN NUMBER OF SAMPLES. SUBSEQUENT ANALYSES OF THE PATIENTS' SAMPLES, ON AN ALTERNATE UNICEL DXC 800 SYSTEM, RECOVERED HIGHER PRE-ALBUMIN RESULTS FOR BOTH PATIENTS. PATIENT (B)(6) OBTAINED AN INITIAL (B)(6) RESULT OF (B)(6) AND A REANALYZED VALUE OF (B)(6). PATIENT (B)(6) OBTAINED AN INITIAL RESULT OF (B)(6) AND A REANALYZED VALUE OF (B)(6). THE ERRONEOUS RESULTS WERE RELEASED OUT OF THE LABORATORY. IT IS UNKNOWN IF PATIENT TREATMENT WAS IMPACTED. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AND VERIFIED THE CARTRIDGE CHEMISTRY SAMPLE SYRINGE TO BE SLIGHTLY LOOSE. THE CUSTOMER STATED QUALITY CONTROL (QC) WAS WITHIN THE ACCEPTABLE RANGE ON THE DAY OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83653 UNICEL® DXC 800 SYNCHRON® SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1