FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2981721 · Received February 27, 2013

Report

Report Number
3004209178-2013-91406
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 14, 2013
Report Date
February 14, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS UNKNOWN. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. THE CUSTOMER STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CALLER TO RUN THE DISPLACEMENT TEST AND PASSED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83795 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 61 YR UNOMEDICAL INFUSION SET