PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-91398
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- February 14, 2013
- Report Date
- February 14, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
THE INSULIN PUMP PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TEST. THE INSULIN PUMP RECEIVED WITH BLEEDING LCD GLASS AND SCRATCHED LCD WINDOW. INTERMITTENT BUTTON PRESS NOTED DURING TESTING DUE TO UNLOCKED LCD CONNECTOR. NO DRIVE MOTOR ANOMALY NOTED.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
IT WAS REPORTED THAT THE CUSTOMER WAS HAVING ISSUES WITH THE INSULIN PUMP. THE BLOOD GLUCOSE READING WAS UNKNOWN. THE CALLER STATED THAT THE BACK END OF THE PISTON WAS COMING OUT, AND THE BUTTONS WERE STICKY. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED.