FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2981714 · Received February 27, 2013

Report

Report Number
3004209178-2013-91398
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 14, 2013
Report Date
February 14, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TEST. THE INSULIN PUMP RECEIVED WITH BLEEDING LCD GLASS AND SCRATCHED LCD WINDOW. INTERMITTENT BUTTON PRESS NOTED DURING TESTING DUE TO UNLOCKED LCD CONNECTOR. NO DRIVE MOTOR ANOMALY NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HAVING ISSUES WITH THE INSULIN PUMP. THE BLOOD GLUCOSE READING WAS UNKNOWN. THE CALLER STATED THAT THE BACK END OF THE PISTON WAS COMING OUT, AND THE BUTTONS WERE STICKY. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84477 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 42 YR