FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2981579 · Received February 27, 2013

Report

Report Number
3008382007-2013-03808
Event Type
Malfunction
Date Received
February 27, 2013
Report Date
January 29, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (03/26/2013)-DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDING: THE TEST STRIP PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

THE LAY USER/PATIENT'S PRODUCT(S) HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED IN THIS CASE HAVE PASSED ALL TESTING AND THE COMPLAINT WAS NOT CONFIRMED. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/PATIENT IN (B)(6) CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WAS READING A BLOOD SAMPLE AS A CONTROL SOLUTION TEST. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84511 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3365800

Patients

Seq Age Sex Outcome Treatment
1