FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 2981490 · Received February 27, 2013

Report

Report Number
1416980-2013-04847
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 7, 2013
Report Date
February 13, 2013
Manufacturer
BAXTER HEALTHCARE - AIBONITO
Product Code
LJH
PMA / PMN Number
K960787
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: ONE ACTUAL DEFECTIVE UNIT WAS AVAILABLE AT THE PLANT FOR EVALUATION. DURING VISUAL INSPECTION WAS OBSERVED A PUNCTURE/ BROKEN BURETTE CHAMBER. PRESSURE TEST WAS PERFORMED AT 8PSI AND THE UNIT LEAKS WHERE THE DEFECT WAS IDENTIFIED DURING VISUAL INSPECTION. FUNCTIONAL TEST WAS ALSO PERFORMED. TIP PROTECTORS WERE REMOVED, CLAMPS WERE CLOSED. UNIT WAS CONNECTED TO A SOLUTION CONTAINER CLAMP WERE OPENED, PROCEED WITH THE PRIMING PROCESS. UNIT FAILED TO FUNCTIONAL TEST DUE A LEAK IN THE BURETTE. ADDITIONAL INFORMATION: SHOULD ADDITIONAL RELEVANT INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A FOLLOW-UP MDR WILL BE SUBMITTED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN RECEIVED FOR EVALUATION. FOLLOWING EVALUATION OR IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LEAK WAS OBSERVED FROM THE DRIP CHAMBER OF A FOUR LEAD TUR IRRIGATION SET DURING OPERATION. THERE WAS A PATIENT INVOLVEMENT BUT NO PATIENT INJURY/ADVERSE EVENT OR MEDICAL INTERVENTION IN ASSOCIATION WITH THE EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84739 ACCESS SYSTEM, IRRIGATION, UROLOGICAL LJH BAXTER HEALTHCARE - AIBONITO UR10D09142