FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2981486 · Received February 27, 2013

Report

Report Number
3015876-2013-00152
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 1, 2013
Report Date
February 1, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K102972
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO THEN REPLACED THE POWER PCB ASSEMBLY AND AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EXAMINED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS THAT PINS 12 TO 13 OF AN INTEGRATED CIRCUIT, DESIGNATOR U5, WERE ELECTRICALLY SHORTED.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO THEN REPLACED THE POWER PCB ASSEMBLY AND AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EXAMINED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS THAT PINS 12 TO 13 OF AN INTEGRATED CIRCUIT, DESIGNATOR U5, WERE ELECTRICALLY SHORTED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NO LONGER POWER ON USING AC OR DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84806 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC

Patients

Seq Age Sex Outcome Treatment
1