ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-02228
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Report Date
- February 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: 04/23/2012 DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: ON EXAMINATION, THE KEYPAD IS DAMAGED; THE KEYPAD COVER HAS BEEN TORN OFF AT THE OK BUTTON WITH THE BUTTON CONTACT EXPOSED. A DAMAGED KEYPAD RUBBER WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD RUBBER SHOULD BE CLEARLY VISIBLE AND PROHIBIT THE USE OF THE KEYPAD BUTTONS. USERS MAY EXPECT KEYPAD DAMAGE DURING NORMAL WEAR AND THE OWNER'S BOOKLET INSTRUCTS THE USER TO CONTACT CUSTOMER SERVICE IF THE USER SUSPECTS THE PUMP MAY BE DAMAGED. ON TESTING, ALL THE KEYPAD BUTTONS DEMONSTRATED INTERMITTENT RESPONSE AND REQUIRED MULTIPLE PRESSES WITH INCREASING FORCE TO EVOKE A RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE BUTTONS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DATE OF SUBMISSION OF FOLLOW UP #1 IS (B)(4) 2013. DATE OF ADDITIONAL ANALYSIS IN FOLLOW UP #1 IS (B)(4) 2013.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGE PRIOR TO DAMAGE) ISSUE. IT WAS REPORTED THAT ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE BUTTON PRESSES TO ELICIT A RESPONSE. REPORTEDLY, THE RUBBER KEYPAD WAS PEELING UP BELOW THE OK BUTTON AND THE TACTILE CHANGES HAD OCCURRED PRIOR TO THE KEYPAD DAMAGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 84442 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |