FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 2981457 · Received February 27, 2013

Report

Report Number
3015876-2013-00151
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
January 29, 2013
Report Date
January 31, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE THERAPY PCB ASSEMBLY WHICH RESOLVED THE REPORTED FAILURE. AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EXAMINED THE REMOVED THERAPY PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS THAT A FILTER, DESIGNATOR FL30, WAS BROKEN AND MEASURED OPEN. THIS FAILURE PREVENTED THE UNIT FROM POWERING ON.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER UP ON AC POWER. UPON EXAMINATION OF THE DEVICE, PHYSIO OBSERVED THAT IT WOULD NOT POWER ON AT ALL (NO AC OR DC POWER). THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.