FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2981321 · Received February 27, 2013

Report

Report Number
1416980-2013-04823
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
January 1, 2013
Report Date
January 15, 2013
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION OF "DID NOT FUNCTION" WAS CONFIRMED DURING DEVICE EVALUATION AS FAILURE CODE 94. THE ASSIGNABLE CAUSE WAS A DAMAGED MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP THAT "DID NOT FUNCTION". UPON FURTHER INVESTIGATION, THE QUALITY ENGINEER CONFIRMED THE REPORTED CONDITION AS A FAILURE CODE 94. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84318 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1