UV FLASH TRANSFER SETS
Report
- Report Number
- 1416980-2013-04818
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- February 6, 2013
- Report Date
- February 7, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. HOWEVER, THE SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THIS COMPLAINT WAS NOT CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED. THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. LEAK TESTING WAS PERFORMED WITH NO ISSUES NOTED. A CLEAR PASSAGE TEST AND CLAMP FUNCTION TEST WERE PERFORMED WITH NO ISSUES NOTED.
IT WAS REPORTED THAT A PATIENT HAD A CONNECTION ISSUE WITH A UV FLASH TRANSFER SET DURING PERITONEAL DIALYSIS (PD) THERAPY. THE REPORTER STATED THAT THE SPIKE CONNECTOR COVER WAS SEPARATED FROM THE SPIKE OF THE TRANSFER SET. THERE WAS PATIENT INVOLVEMENT BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83745 | UV FLASH TRANSFER SETS | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |