FDA Adverse Event Malfunction Summary report: N

PD CYCLER 115V, HOMECHOICE

MDR report key: 2981098 · Received February 27, 2013

Report

Report Number
1416980-2013-04793
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 7, 2013
Report Date
February 7, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED ISSUE WAS CONFIRMED OVER THE PHONE. THE DEVICE PASSED BOTH THE HC RETURN INSTRUMENT TEST / EVALUATION (RITE) ELECTRICAL TEST AND THE RITE FUNCTIONAL TEST. REVIEW OF SERVICE HISTORY REVEALED NO FAILURES/PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY AND THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. REVIEW OF THE DEVICE SERVICE HISTORY REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY OF IIPV-ADULT. THE ASSIGNABLE CAUSE WAS UNDETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HIGH DRAIN 105 ALARM OCCURRED DURING THERAPY ON THE HOMECHOICE MACHINE(HC). THE HOME PATIENT(HP) STATED HE HAD NO PROBLEMS DURING THERAPY LAST NIGHT AND WANTED TO CONTINUE THERAPY. THE HC UNIT WAS OPERATIONAL. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83435 PD CYCLER 115V, HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE AUTOMATED PD SET WITH CASSETTE