FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 2981066 · Received February 19, 2013

Report

Report Number
1028232-2013-00339
Event Type
Injury
Date Received
February 19, 2013
Date of Event
November 6, 2012
Report Date
February 6, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
DTB
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THERE WAS A LEAD DISLODGEMENT ABOUT THREE WEEKS AGO. THE LEAD WAS SUCCESSFULLY REPOSITIONED. THE PT NOW PRESENTS WITH PERICARDIAL EFFUSION. THE PHYSICIAN WAS SCHEDULED TO DRAIN THE EFFUSION. THERE WERE NO ADDITIONAL ADVERSE PT EFFECTS. ALL AVAILABLE INFORMATION INDICATES THE DEVICE REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71428 DEXTRUS 4136 PACER LEAD DTB BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization