FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION
MDR report key: 2981038
·
Received January 29, 2013
Report
- Report Number
- 2221819-2013-00046
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- November 12, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 011540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVE EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENTS OF THE SYSTEM HARD DRIVES. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATION. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A CORRUPTED HARD DRIVE ON THE PANORAMA CENTRAL STATION DURING BOOT UP, WHICH MAY HAVE RESULTED IN LOSS OF AMBULATORY TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38524 | PANORAMA CENTRAL STATION | NONE | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|