FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 2980999 · Received January 29, 2013

Report

Report Number
2221819-2013-00034
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
October 8, 2012
Report Date
January 30, 2013
Manufacturer
MINDRAY DS USA, INC
Product Code
MHX
PMA / PMN Number
011540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EVALUATED THE SYSTEM AND FOUND A FAILED MOTHER BOARD. CORRECTIONS INCLUDED REPLACEMENT OF THE SYSTEM'S MOTHER BOARD AND CLEARING OF ERROR LOGS. THE UNIT WAS TESTED TO FACTORY'S SPECS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE PANORAMA CENTRAL STATION HAD A BLANK SCREEN, WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39767 PANORAMA CENTRAL STATION NONE MHX MINDRAY DS USA, INC

Patients

Seq Age Sex Outcome Treatment
1