FDA Adverse Event
Malfunction
Summary report: N
PANORAMA SERVER
MDR report key: 2980997
·
Received January 29, 2013
Report
- Report Number
- 2221819-2013-00036
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- October 4, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MINDRAY DS USA, INC
- Product Code
- MHX
- PMA / PMN Number
- 031760
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EVALUATED THE SYSTEM AND VERIFIED THE PROBLEM. CORRECTIONS INCLUDED CLEARING OF THE SYSTEM'S ERROR LOGS AND REPLACEMENTS OF THE PANORAMA WIRELESS TRANSCEIVER (TIM) AND RADIO TRANSCEIVER BOARD. THE SYSTEM WAS TESTED TO FACTORY'S SPECS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED PANORAMA CENTRAL STATION HAD LOST COMMUNICATION ON ALL CHANNELS, WHICH MAY HAVE RESULTED IN LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38720 | PANORAMA SERVER | NONE | MHX | MINDRAY DS USA, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |