FDA Adverse Event Injury Summary report: N

SETROX S 45

MDR report key: 2980991 · Received February 19, 2013

Report

Report Number
1028232-2013-00448
Event Type
Injury
Date Received
February 19, 2013
Date of Event
September 24, 2012
Report Date
February 5, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
JXK
PMA / PMN Number
P950038
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

THIS SYSTEM WAS REMOVED DUE TO (B)(6) ENDOCARDITIS ON (B)(6) 2012. THERE IS NO INDICATION THE SYSTEM WAS REPLACED. SHOULD ADDITIONAL INFO BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71157 SETROX S 45 PACER LEAD JXK BIOTRONIK SE & CO. KG 350973

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization