FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION
MDR report key: 2980990
·
Received January 29, 2013
Report
- Report Number
- 2221819-2013-00018
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- September 24, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 011540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE POWER SUPPLY AND REPROGRAMMING OF THE WIRELESS CARD. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED WHILE PTS WERE BEING MONITORED ON THE PANORAMA CENTRAL STATION WITH AMBULATORY TELEPACKS, THE CENTRAL STATION HAD LOST COMMUNICATION. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 39842 | PANORAMA CENTRAL STATION | NONE | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |