FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 2980990 · Received January 29, 2013

Report

Report Number
2221819-2013-00018
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
September 24, 2012
Report Date
January 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
011540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE POWER SUPPLY AND REPROGRAMMING OF THE WIRELESS CARD. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED WHILE PTS WERE BEING MONITORED ON THE PANORAMA CENTRAL STATION WITH AMBULATORY TELEPACKS, THE CENTRAL STATION HAD LOST COMMUNICATION. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39842 PANORAMA CENTRAL STATION NONE MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1