FDA Adverse Event
Malfunction
Summary report: N
PANORAMA
MDR report key: 2980985
·
Received January 29, 2013
Report
- Report Number
- 2221819-2013-00021
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- August 16, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 011540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENTS OF THE SYSTEM HARD DRIVE AND POWER SUPPLY. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE PANORAMA CENTRAL STATION SHUT DOWN AND HAD A BLANK SCREEN THAT DISPLAYED AN ERROR MESSAGE, WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38718 | PANORAMA | NONE | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |