FDA Adverse Event Malfunction Summary report: N

PANORAMA

MDR report key: 2980985 · Received January 29, 2013

Report

Report Number
2221819-2013-00021
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
August 16, 2012
Report Date
January 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
011540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENTS OF THE SYSTEM HARD DRIVE AND POWER SUPPLY. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE PANORAMA CENTRAL STATION SHUT DOWN AND HAD A BLANK SCREEN THAT DISPLAYED AN ERROR MESSAGE, WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
38718 PANORAMA NONE MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1