FDA Adverse Event Malfunction Summary report: N

CADIERE FORCEPS INSTRUMENT

MDR report key: 2980871 · Received February 27, 2013

Report

Report Number
2955842-2013-00647
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
December 3, 2012
Report Date
January 29, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE NONCONFORMANCE WAS A BROKEN GRIP CABLE. THE INTUITIVE MOTION WAS NOT BEING EXPERIENCED BECAUSE THE GRIP CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND EXHIBITED RIM PULLEY DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED PRIOR TO STARTING A DA VINCI S SURGICAL PROCEDURE THAT THE CADIERE FORCEPS INSTRUMENT TIP DID NOT WORK. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84597 CADIERE FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420049-06 M10120229 142

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI S SYSTEM, INSTRUMENTS & ACCESSORIES