DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00057
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- January 27, 2013
- Report Date
- January 31, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- KXT
- PMA / PMN Number
- K093441
- Removal / Correction Number
- 2517506-03-06-2013-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS CONFIRMED CUSTOMER COMPLAINTS OF LOW QC AND PATIENT RECOVERY FOR DIGOXIN WITH THIS CALIBRATOR LOT. INTERNAL TESTING HAS CONFIRMED AN AVERAGE 15% LOW BIAS IN THE THERAPEUTIC RANGE OF 0.9 TO 2.0 NG/ML. THE CAUSE OF THE BIAS LOW IN DIGOXIN RECOVERIES IS CALIBRATOR INACCURACY. SIEMENS HEALTHCARE DIAGNOSTICS INC. CONDUCTED A VOLUNTARY RECALL FOR DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053. AN URGENT MEDICAL DEVICE RECALL, COMMUNICATION #(B)(4), WAS ISSUED IN (B)(4) 2013 TO IMPACTED CUSTOMERS. CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE AND TO DISCARD ANY REMAINING INVENTORY OF DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053.
(B)(4).
BIASED LOW DIGOXIN RESULTS WERE OBTAINED ON QC SAMPLES UPON CALIBRATION WITH A NEW LOT OF CALIBRATOR. PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED AS A RESULT OF THE BIASED LOW DIGOXIN RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE BIASED LOW DIGOXIN RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85021 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | DRUG 4 CALIBRATOR | KXT | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 2KD052 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |