FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION
MDR report key: 2980746
·
Received January 29, 2013
Report
- Report Number
- 2221819-2013-00014
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- September 18, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MINDRAY DS USA, INCL
- Product Code
- MHX
- PMA / PMN Number
- 03760
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REBOOTING, CLEARING OF ERROR LOGS, AND REPLACEMENT OF THE SYSTEM HARD DRIVERS. THE SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE PANORAMA CENTRAL STATION OPERATED SLOWLY, HAD AN ORANGE SCREEN AND THEN CRASHED, WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 39934 | PANORAMA CENTRAL STATION | NONE | MHX | MINDRAY DS USA, INCL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|