FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2980732 · Received February 27, 2013

Report

Report Number
2531779-2013-02201
Event Type
Malfunction
Date Received
February 27, 2013
Report Date
January 30, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED LOSS OF PRIME WITH ZERO FORCE IN THE BLACK BOX HISTORY. THE REWIND, PRIME AND LOAD STEPS WERE PERFORMED ON THE PUMP WITH NO LOSS OF PRIME ISSUES WITH LOSS OF PRIME. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL PROGRAM WITH NO LOSS OF PRIME. INVESTIGATION REVEALED THAT THE FORCE SENSOR FLEX PINS WERE FOUND TO BE PARTIALLY DISLODGED FROM THE PRINTED CIRCUIT BOARD (PBB). ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE FORCE SENSOR FLEX PINS WERE FOUND TO BE PARTIALLY DISLODGED FROM THE PRINTED CIRCUIT BOARD (PBB). THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84886 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR