DIGITAL RETINAL CAMERA CF-1
Report
- Report Number
- 1000181430-2013-00018
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Report Date
- September 26, 2011
- Manufacturer
- CANON INC.
- Product Code
- HKI
- PMA / PMN Number
- K063717
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS MDR IS BEING SUBMITTED RETROACTIVELY AS A RESULT OF AN INTERNAL AUDIT OF COMPLAINT FILES CONDUCTED BY CANON HEALTHCARE SOLUTIONS USA. GENDER INFORMATION WAS NOT PROVIDED. (B)(4). THE SERVICE REPRESENTATIVE REPLACED THE SPLIT PRISM MOTOR. HE ALIGNED THE SPLIT PRISM AND ADJUSTED TO OD POSITION. HE CLEANED THE OPTIC AND LUBRICATED THE QUICK RELEASE MIRROR. THE SYSTEM WAS THEN TESTED. (B)(4).
THE CUSTOMER REPORTED THAT THERE WAS A SPLIT LINE SHADOW IN THE IMAGES. NO INFORMATION WAS PROVIDED REGARDING WHAT IMAGE TYPES (COLOR, RED-FREE, FLUORESCEIN ANGIOGRAPHY) WERE AFFECTED. THE CUSTOMER DID NOT STATE THAT THE FLUORESCEIN IMAGING WAS REPEATED. HOWEVER, IF THE PROBLEM OCCURRED DURING THE FLUORESCEIN IMAGING PORTION OF THE EXAM, THE TECHNICIAN MAY HAVE RESCHEDULED THE EXAM. A RESCHEDULED EXAM WOULD HAVE INVOLVE A REPEAT INJECTION OF FLUORESCEIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38224 | DIGITAL RETINAL CAMERA CF-1 | HKI | CANON INC. | CF-1 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |