FDA Adverse Event Malfunction Summary report: N

DIGITAL RETINAL CAMERA CX-1

MDR report key: 2980369 · Received January 28, 2013

Report

Report Number
1000181430-2013-00028
Event Type
Malfunction
Date Received
January 28, 2013
Report Date
May 16, 2012
Manufacturer
CANON INC.
Product Code
NFJ
PMA / PMN Number
K092565
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED RETROACTIVELY AS A RESULT OF AN INTERNAL AUDIT OF COMPLAINT FILES CONDUCTED BY CANON HEALTHCARE SOLUTIONS USA. (B)(4). THE SERVICE ENGINEER REPLACED THE SPLIT PRISM ASSEMBLY. HE ADJUSTED THE OD CALIBRATION AND CHECKED THE OPTICAL SYSTEM. THE SYSTEM WAS TESTED FOR PERFORMANCE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WERE CIRCULAR SHADOWS ON THE LEFT SIDE OF THE IMAGES. NO INFORMATION WAS PROVIDED REGARDING WHAT IMAGE TYPES (COLOR, RED-FREE, FLUORESCEIN ANGIOGRAPHY) WERE AFFECTED. THE CUSTOMER DID NOT STATE THAT THE FLUORESCEIN IMAGING WAS REPEATED. HOWEVER, IF THE PROBLEM OCCURRED DURING THE FLUORESCEIN IMAGING PORTION OF THE EXAM, THE TECHNICIAN MAY HAVE RESCHEDULED THE EXAM. A RESCHEDULED EXAM WOULD HAVE INVOLVED A REPEAT INJECTION OF FLUORESCEIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37306 DIGITAL RETINAL CAMERA CX-1 NFJ CANON INC. CX-1 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK