FDA Adverse Event
Malfunction
Summary report: N
MERIT CUSTOM KIT
MDR report key: 2980176
·
Received February 21, 2013
Report
- Report Number
- 1721504-2013-00040
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- January 28, 2013
- Report Date
- January 28, 2013
- Manufacturer
- MERIT MEDICAL SYSTEMS, INC.
- Product Code
- DQO
- PMA / PMN Number
- K994253
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: ONE NEW UNUSED DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NO EXCEPTION DOCUMENTS. THE COMPLAINT DATABASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE DEVICE WAS EXAMINED VISUALLY, PARTICULATE LARGER THAN THE ACCEPTANCE CRITERIA WAS FOUND. THE COMPLAINT IS CONFIRMED FOR THIS DEVICE. THE ROOT CAUSE IS ATTRIBUTED TO THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
THE DISTRIBUTOR REPORTED CONTAMINATION WAS IDENTIFIED IN THE BARREL OF THE SYRINGE WITHIN THE KIT DURING THEIR INITIAL INSPECTION OF RECEIVED PRODUCT. THE DEVICE WAS NOT SENT TO A USER FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76975 | MERIT CUSTOM KIT | CATHETER, INTRAVASCULAR, DIAGNOSTIC | DQO | MERIT MEDICAL SYSTEMS, INC. | H404008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |