FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 2979817 · Received February 27, 2013

Report

Report Number
9611451-2013-00147
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
December 26, 2012
Report Date
January 30, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED THAT THE SPIKE AND THE FEEDSET TUBING WERE RETURNED SEPARATED FROM EACH OTHER. SUFFICIENT GLUE WAS PRESENT AROUND THE SPIKE TUBING CONNECTION; HOWEVER THE GLUE HAD NOT BONDED PROPERLY. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120504. CONCLUSION: THE SEPARATION OF THE SPIKE AND FEEDSET TUBE WAS DUE TO THE FAILURE OF THE GLUE BOND. WE WERE UNABLE TO DETERMINE THE CAUSE FOR THIS. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE FAILURE OF THE BOND DEVELOPED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT WATER WAS LEAKING FROM THE CONNECTION BETWEEN THE WATER BAG AND THE SPIKE ON THE WATER FEEDSET TUBE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND AFTER SEVEN DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84580 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 120504

Patients

Seq Age Sex Outcome Treatment
1