FDA Adverse Event Malfunction Summary report: N

COROMETRICS

MDR report key: 297738 · Received September 22, 2000

Report

Report Number
MW1020004
Event Type
Malfunction
Date Received
September 22, 2000
Report Date
September 22, 2000
Manufacturer
GE MEDICAL SYSTEMS/MARQUETTE
Product Code
KXN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

ADD'L INFO REC'D FROM MFR 11/13/00: IN PERFORMING A RISK ASSESSMENT, IT WAS DETERMINED THAT, IN ORDER, FOR DATA TO BE SENT TO THE WRONG BED, THREE SIMULTANEOUS EVENTS MUST OCCUR -(1) THE MONITORS MUST BE ON A COROLAN NETWORK; (2) ADDRESS CHECKING IN THE MONITORS / WALLPLATES IS NOT TURNED ON; AND (3) A HARDWARE FAILURE IN THE COROLAN OR THE CABLES MUST OCCUR. SINCE THE LIKELIHOOD OF ALL THREE EVENTS OCCURRING SIMULTANEOUSLY IS SMALL, THE PROBABILITY OF OCCURRENCE IS NEGLIGIBLE. AS DISCUSSED ABOVE, THE HOSP QS SYSTEM HAS BEEN INSTALLED FOR OVER 5 YEARS AND PREVIOUSLY HAD NOT REPORTED PROBLEMS OF THIS NATURE. IT IS LIKELY THAT STRESSES WERE PLACED ON THE CABLES OR MONITORS, CAUSING A HARDWARE FAILURE (OPEN ADDRESS LINES). MFR'S INFO TECHNOLOGIES' SVC WENT TO THE FACILITY TO VERIFY THAT THE MONITORS, CABLES, WALLPLATES, ETC. WERE WORKING PROPERLY WHILE THE MONITOR AND COROLAN NETWORK CONTINUED TO BE USED AT HOSP. NO PROBLEMS WERE FOUND. IN RESPONSE TO THE CUSTOMER REQUEST, SVC HAS COMPLETED THE PROCESS OF REPLACING THE COROLAN NETWORK WITH A SERIAL MONITORING NETWORK.

Description of Event or Problem · 1

DATA FROM THE FETAL MONITORS IS BEING INCORRECTLY PLACED IN THE ELECTRONIC CHARTING SYSTEM. DATA IS BEING TRANSPOSED FROM ONE PT TO ANOTHER - E.G., BED #1 DATA MAY BE INCORRECTLY ASSIGNED TO BED #2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COROMETRICS FETAL MONITOR KXN GE MEDICAL SYSTEMS/MARQUETTE COROMETRICS 118 *
2 QMI ELECTRONIC CHARTING DSF GE MEDICAL SYSTEMS/MARQUETTE QMI *

Patients

Seq Age Sex Outcome Treatment
1 * Other