FDA Adverse Event Malfunction Summary report: N

75MM SELECTABLE NEW TLC

MDR report key: 2974842 · Received February 22, 2013

Report

Report Number
3005075853-2013-00816
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
February 7, 2013
Report Date
February 8, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K092577
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD HAD THE PROXIMAL 25 DRIVERS UP WITHOUT STAPLES, AND THE REMAINING DRIVERS DOWN WITH STAPLES PRESENT. THE RETURNED CARTRIDGE RELOAD HAD THE SWING TAB IN THE LOCKED POSITION, WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. IT SHOULD BE NOTED THAT THE CARTRIDGE RELOAD IS DESIGNED TO LOCKOUT, AS A SAFETY FEATURE, IF ANY STAPLES HAVE BEEN FIRED FROM THE CARTRIDGE RELOAD. IN ADDITION FAILURE TO COMPLETE THE STROKE MAY RESULT IN INCOMPLETE STAPLE LINE. THE CARTRIDGE RELOAD WAS MANUALLY UNLOCKED AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED CARTRIDGE RELOAD AND IT FIRED, CUT, AND FORMED ALL THE REMAINING STAPLES AS INTENDED. THE DEVICE FIRED WITH NO DIFFICULTIES AND THE STAPLES WERE NOTED TO HAVE THE PROPER SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLOSTOMY TAKEDOWN PROCEDURE THE SURGEON ATTEMPTED TO FIRE THE DEVICE FOR THE FIRST TIME ACROSS THE COLON AND IT JAMMED AND WOULD NOT FIRE. THEY RELOADED THE DEVICE AND WHEN THEY FIRED IT AFTER THEY OBSERVED THE STAPLE LINE THE STAPLES WERE MALFORMED. THE CASE WAS COMPLETED WITH A THIRD DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78410 75MM SELECTABLE NEW TLC STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK H44144

Patients

Seq Age Sex Outcome Treatment
1 RELOAD - SR75