FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2974776 · Received January 17, 2013

Report

Report Number
1811755-2013-90094
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 19, 2012
Report Date
December 19, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVAL, CORROSION WAS FOUND IN THE SOCKETS, WHICH IS A PROBABLE CAUSE OF THE DEVICE RUNNING WITHOUT USER ACTIVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE SAW WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27309 MICRO SAGITTAL SAW ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK