MICRO SAGITTAL SAW
MDR report key: 2974776
·
Received January 17, 2013
Report
- Report Number
- 1811755-2013-90094
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Date of Event
- December 19, 2012
- Report Date
- December 19, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVAL, CORROSION WAS FOUND IN THE SOCKETS, WHICH IS A PROBABLE CAUSE OF THE DEVICE RUNNING WITHOUT USER ACTIVATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE SAW WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.