FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2974733 · Received January 18, 2013

Report

Report Number
1824206-2013-00559
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
March 15, 2012
Report Date
March 15, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE HYDRAULIC MANIFOLD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECH ALLEGED THE HEAD HI/LOW WAS DRIFTING DOWN SLOWLY. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28769 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1