FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2974607 · Received January 17, 2013

Report

Report Number
1811755-2013-90082
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 17, 2012
Report Date
December 18, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, CORROSION WAS FOUND IN THE INTERNAL COMPONENTS, INCLUDING THE ROTOR, MOTOR, DIRVESHAFT ASSEMBLY AND BEARINGS. THE INCREASED FRICTION OF THE CORRODED PARTS IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING. THE REASON FOR THE SERIALIZATION STARTING WITH 90XXX IS TO PROVIDE CLARIFICATION FOLLOWING A CHANG E IN STRYKER'S ELECTRONIC COMPLAINT SYSTEMS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO PT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27319 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK