FDA Adverse Event Malfunction Summary report: N

PNEUMATIC ROTO OSTEOTOME DRILL

MDR report key: 2974604 · Received January 17, 2013

Report

Report Number
1811755-2013-90079
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 17, 2012
Report Date
December 17, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
GET
PMA / PMN Number
K863447
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE MOTOR CARTRIDGE ASSEMBLY AND DRIVESHAFT REQUIRED REPLACEMENT DUE TO THE OVERHEATING OF THE DEVICE, WHICH WAS LIKELY CAUSED BY USE OVER TIME. THE AFFECTED COMPONENTS WERE REPLACED. THE DRILL WAS CLEANED, LUBED, ADJUSTED AND RETURNED TO THE CUSTOMER. THE REASON FOR THE SERIALIZATION STARTING WITH (B)(4) IS TO PROVIDE CLARIFICATION FOLLOWING A CHANGE IN STRYKER'S ELECTRONIC COMPLAINT SYSTEMS.

Description of Event or Problem · 1

THE PNEUMATIC ROTO OSTEOTOME DRILL WAS RETURNED FOR SVC. UPON EVALUATION AT THE MFR, IT WAS FOUND TO OVERHEAT. THERE WAS NO PT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27318 PNEUMATIC ROTO OSTEOTOME DRILL GET STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK