Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) HAS BEEN CONFIRMED. UPON INVESTIGATION THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND DN. THE ROOT CAUSE OF THE DAMAGED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULT FROM THE DAMAGE ELECTRODE BELT CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PATIENT'S DAUGHTER CALLED ZOLL CUSTOMER SUPPORT TO REPORT DAMAGED CABLE OR WIRES EXPOSED ON THE PATIENT'S ELECTRODE BELT. THE PATENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.