FDA Adverse Event
Malfunction
Summary report: N
CURLIN INFUSION IV PUMP
MDR report key: 2974566
·
Received January 29, 2013
Report
- Report Number
- 1722139-2013-00117
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- January 22, 2013
- Report Date
- January 22, 2013
- Manufacturer
- MOOG MED DEVICES GROUP
- Product Code
- FRN
- PMA / PMN Number
- K981816
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FOUND THAT PUMP SHOWED IMPACT BENT PLATEN CAUSING FREE FLOW AND PUMP FAILED 40 PSI TEST. PLATEN WAS REPLACED TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THAT PUMP CONSTANTLY FREE FLOWS WHEN THE SET ATTACHES DURING TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 39792 | CURLIN INFUSION IV PUMP | FRN | MOOG MED DEVICES GROUP | 4000 CMS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |