FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2974389 · Received February 22, 2013

Report

Report Number
1416980-2013-04479
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
November 12, 2012
Report Date
November 12, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE DAMAGED FORCE SENSING RESISTORS (FSRS) WAS DETERMINED TO BE DUE TO USE OF ALCOHOL IN THE CHANNEL. TO CORRECT THE CONDITION, THE FSRS WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICING, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE DAMAGED FORCE SENSING RESISTORS (FSRS). NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77792 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1