ROD HOLDING FORCEPS FOR 6.0MM RODS
Report
- Report Number
- 8030965-2013-00586
- Event Type
- Malfunction
- Date Received
- February 22, 2013
- Date of Event
- January 24, 2013
- Report Date
- January 24, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- HTD
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- NURSE
Narratives
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE FORCEPS WERE RECEIVED WITH SOME MINOR SCRATCHES ON THE SURFACE. THE SPIKES LOCATED IN THE FUNCTIONAL END OF BOTH FORCEPS MAY BE WORN FROM USE. BOTH FORCEPS ARE FROM LOTS THAT WERE MANUFACTURED OVER 10 YEARS AGO. CHU ENGINEER WAS UNABLE TO PASS A 4.99 AND 5.01 GAGE PIN THRU THE FORCEPS WHICH CORRESPONDS TO THE DESIGN SPECIFICATION ON THE DRAWING. A 5MM MAXIMUM DISTANCE BETWEEN THE SPIKES IS CALLED OUT ON THE DRAWING FOR THESE FORCEPS. IN ADDITION, THE COMPLAINT INVOLVES USING THESE WITH THE MATRIX SYSTEM AND THESE ARE NOT TO BE USED WITH MATRIX.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. THE DEVICE HISTORY RECORD IS NOT AVAILABLE ANYMORE FOR REVIEW AS THE DEVICE IS OVER 11 YEARS OLD. THEREFORE WE CONCLUDE THIS COMPLAINT TO BE INDETERMINATE.
ON (B)(6) 2013 A SURGEON REPORTED WHILE USING ROD HOLDING FORCEPS TO PERFORM ROTATION OF THE ROD; THE ROD HOLDERS WOULD SLIP ON THE ROD AND NOT GRIP HARD ENOUGH. THE SLIPPAGE WOULD NOT ALLOW HIM TO ROTATE THE ROD TO HIS SATISFACTION, SO MAXIMUM CORRECTION WAS NOT ACHIEVED. THIS ADDED 15 MINUTES TO THE SURGERY AND THE SURGERY WAS COMPLETED WITHOUT FURTHER INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78904 | ROD HOLDING FORCEPS FOR 6.0MM RODS | FORCEPS | HTD | SYNTHES GMBH | A7KA20 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |